Medical Device Sterilization Validation Tests

Photo

In order to release your medical device among the health-care community, it must be able to be sterilized. You also need a team of qualified and knowledgeable medical device sterility test validation professionals to determine that the method is efficacious. HIGHPOWER’s medical device sterilization validation methodology is based on methods outlined in AAMI TIR 12, ANSI / AAMI / ISO 14937, ANSI / AAMI ST81, and ANSI/AAMI/ISO 17665-1. Reusable medical devices require sterilization efficacy validation in order to assure proper and safe reprocessing by health care personnel. After that method is validated, it can be repeated per the IFU after each use.

Sterilization Efficacy testing is sometimes referred to as “Half Cycle Testing, 100% Overkill Method Testing or Sterilization Validation Testing.” All these terms mean the same thing in order to determine how effectively a sterilization process will sterilize the device(s) being tested. With the proper medical device efficacy testing, there will be no question that your product is safe to use on patients.

Photo

Typically, the device being validated is run in an FDA-cleared sterilization cycle using one half of the sterilization exposure time that is normally used in a clinical setting. The test device also includes at least 10^6 spores of a highly resistant organism, placed in locations that would be the most difficult for sterilant penetration. A full cycle then result with at least 10-6 SAL (sterility assurance level). By testing with such stringent guidelines, we can validate that the sterilization process will be successful even in worst-case testing conditions.

Photo

HIGHPOWER has every major FDA cleared sterilization process onsite. This includes steam sterilization validation modalities as well as low-temperature sterilization processes.

Sterilization is a crucial aspect of the medical device validation process, and there are a number of ways to test sterilization efficacy. We’re constantly adding new medical device sterilization validation tests to meet our customer’s expectations. If you don’t see your required test listed, please contact us at clientservices@highpowervtls.com with your inquiry.

Tests Available by Test Code:

Test Code Test Name
SE-100NX-DUOSterilization Efficacy Validation for STERRAD 100NX Duo
SE-100NX-EXPSterilization Efficacy Validation for STERRAD 100NX Express
SE-100NX-FLXSterilization Efficacy Validation for STERRAD 100NX Flex
SE-100NX-STDSterilization Efficacy Validation for STERRAD 100 NX Standard
SE-100SSterilization Efficacy Validation for STERRAD 100S Cycles
SE-COLDChemistry Immersion Efficacy (High Level Disinfectant/Sterilant)
SE-DRYSterilization Efficacy Validation for Dry Heat
SE-ETOSterilization Efficacy Validation for EtO
SE-NX-ADVSterilization Efficacy Validation for STERRAD NX Advanced
SE-NX-STDSterilization Efficacy Validation for STERRAD NX Standard
SE-STEAMSterilization Efficacy Validation for Steam (gravity or dynamic air removal)
SE-VPROSterilization Efficacy Validation for VPRO processes
TP-STEAMSteam Thermal Profile Study
DT-STEAMSteam Dry Time Study
TP-DRYDry Heat Thermal Profile Study
TP-WASHThermal Disinfection—Ao Study
HLDHigh Level Disinfection Study
ILDIntermediate Level Disinfection
LLDLow Level Disinfection
SIM-USESimulated Use Testing
IN-USEIn Use Testing